With SpO2 (Nellcor optional) and Respironics CO2 highly advanced monitoring technologies, EDAN M3B ensures effective capnography monitoring for intubated and non-intubated patients for continuous long-term monitoring. It is tailored for mechanically ventilated and non-intubated patients.

Features:

  • 5.7 inch high resolution display for easy reading
  • Lightweight, portable design and user-friendly interface for easy operation
  • SpO2, Pulse Rate and Respironics sidestream and mainstream CO2 measurement
  • Nellcor OxiMax SpO2 selectable
  • Display numeric and waveform information simultaneously
  • Nurse call
  • Powerful storage capacity
  • Built-in Lithium-ion Battery for 8 hours working
  • Suitable for adult, pediatric and neonatal patients
  • Trend table review and record
  • Trend graph review and record
  • USB data storage and review
  • Wired and wireless network capability

Configuration:

  • M3B (SpO2 + CO2

Technical Specifications:

Physical Specifications

  • Size: 200.8mm (L) x 241mm (H) x 189mm (D)
  • Weight: 3 kg
  • Display: 5.6” color TFT resolution: 640X480

Power Supply

  • AC Power 100-240 VAC, 50/60HZ
  • Battery Type: Lithium-ion
  • Voltage 14.8 V DC
  • Capacitance: 2,200mAh/4,400 mAh
  • Working Period: 240 min/480 min
  • Rechargeable Period: 150 min / 300 min

Thermal Recorder (Optional)

  • Paper Width: 48 mm
  • Paper Speed: 25 mm/s

SpO2

  • Measurement & Alarm Range: 0~100% (EDAN SPO2)
  • Resolution: 1%
  • Accuracy: +2% (70~100%, Adult/Pediatric) +3% (70~100%, Neonate)
  • PR Measurement and Alarm Range: 25~300bpm
  • Resolution: 1bpm
  • Accuracy: 3bpm
  • Refresh rate: 1s

Nellcor OximaxTM SpO2 (Optional)

  • Measurement & Alarm Range: 0 - 100
  • Resolution: 1%
  • Accuracy: +2-3% (70~100%, Adult/Pediatric) +3.3.5% (70~100%, Neonate)
  • PR Measurement and Alarm Range: 20 - 300 bpm
  • Resolution: 1 bpm
  • Accuracy: 3bpm (depends on probe)
  • Refresh: 1s

Respironics Mainstream CO2

  • CO2 Measurement Range: 0 to 150 mm Hg, 0 to 19.7, 0 to 20 kPa (at 760 mm Hg)
  • Rise Time: Less than 60 ms - Adult/infant Reusable or Single Parent Use Airway Adapter
  • CO2 Resolution: 0.1 mm Hg 0 to 69 mm Hg 0.25 mm Hg 70 to 150 mm Hg
  • CO2 Accuracy: 0-40 mm Hg +2 mm Hg 41-70 mm Hg +5% of reading 71-100 mm Hg +8% of reading 101-150 mm Hg +10% of reading
  • Above 80bpm: +12% of reading

Respironics Sidestream CO2

  • Sample Rate: 50mL/min
  • CO2 Measurement Range: Sidestream
  • CO2 Measurement Range: 0 to 150 mm Hg, 0 to 19.7%, 0 to 20 kPa (at 760 mm Hg) Barometric Pressure supplied by Host
  • CO2 Resolution: 0.1 mm Hg 0 to 69 mm Hg 0.25 mm Hg 70 to 150 mm Hg
  • CO2 Accuracy: 0-40 mm Hg +12 mm Hg 41-70 mm Hg +5% of reading 71-100 mm Hg +8% of reading 101-150 mm Hg +10% of reading
  • Above 80 bpm: +12% of reading
  • Respiratory Rate Range: 2 to 150 bpm
  • Respiratory Rate Accuracy: +1 breath